Certificate of Analysis (CoA)
Usually reports tests and results for an identified sample or batch against acceptance criteria. It may support a material-quality decision only within the applicable GMP system and supplier-control procedure.
AUPAM · Certificate evidence checker
Review the evidence inside a Certificate of Analysis or Certificate of Conformity before relying on its title.
Neither document is automatically more comprehensive. A CoA normally answers a batch-specific analytical question; a CoC attests conformity to a stated scope. The issuer, scope, batch link and underlying evidence determine what either certificate can support.
Usually reports tests and results for an identified sample or batch against acceptance criteria. It may support a material-quality decision only within the applicable GMP system and supplier-control procedure.
Attests that defined requirements were met. Its scope may concern a batch, product, system, standard or market requirement. The title alone does not show which evidence was reviewed.
Select only what you can verify in the document or controlled supporting record. The result reviews coverage; it is not an authority release, disposition or acceptance decision.
Choose the certificate type and mark the evidence you can verify.
Decision-based practice
Fictional training cases. Decide what to do before revealing the suggested path, then apply the national and site procedures.
Scenario: A CoC states that the batch conforms, while the attached CoA shows an assay result outside the approved limit.
Place the material or batch on hold. Verify the specification, result, version and document linkage; open the applicable discrepancy or laboratory investigation. A conformity statement does not cancel a conflicting result.
Authenticated CoA and CoC; approved specification; original analytical record; sampling, method and investigation records; Quality disposition.
Quality Unit; include Regulatory Affairs if a submitted dossier, import record or authority commitment may be affected.
Do not select the more favourable document, average conflicting values, or release only because the CoC says compliant.
Scenario: The CoA includes results and limits, but supplier qualification is incomplete and no receiving identity test is documented.
Keep the material under the applicable receipt status. Follow the approved sampling and identity-testing procedure and complete or escalate supplier qualification before relying on supplier results.
Approved-supplier status; qualification and periodic-review records; receipt, sampling and identity-test records; CoA verification; current quality agreement.
Quality Unit and supplier-quality owner. Procurement may support records but must not make the quality disposition.
Do not treat a polished certificate, prior purchases or commercial urgency as supplier qualification.
Scenario: The batch passes the CoA limits, but they belong to a superseded specification; the approved receiving specification is newer.
Hold disposition while Quality confirms the specification applicable to this batch, date and market. Reconcile change-control implementation and assess affected documents and batches.
Approved specifications and effective dates; change control; purchase requirement; supplier notification; quality agreement; manufacture date; market commitments.
Quality Unit and change-control owner; Regulatory Affairs where registered limits or implementation commitments differ.
Do not compare with whichever limit produces a pass or edit the certificate locally.
Scenario: A valid manufacturing-site GMP certificate is submitted instead of batch-specific analytical evidence for an incoming material.
Record what the GMP certificate covers, then obtain the batch evidence required by the receipt and testing procedure. Keep site-compliance status separate from batch disposition.
GMP certificate scope, issuer and validity; batch CoA or required tests; supplier/manufacturer identity; receipt procedure; approved specification.
Quality Unit; Regulatory Affairs if the certificate is also required for licensing or import documentation.
Do not infer that a GMP-certified site makes every batch compliant, or that a CoA proves the site’s GMP status.
Scenario: The email and supplier portal provide CoAs with the same batch and issue date, but one result and the approval signature differ.
Stop reliance on both copies until authenticity and version history are resolved through a controlled supplier contact. Preserve both and initiate the applicable data-integrity or supplier discrepancy process.
Both original files and metadata; portal audit trail; controlled correspondence; supplier version history; original result; authorised approval record.
Quality Unit, data-integrity lead and supplier-quality owner; Regulatory Affairs if a conflicting copy was submitted externally.
Do not delete the inconvenient copy, rename one final without evidence, or ask for an undocumented replacement.
Scenario: The CoA supports internal quality review, but the current receiving procedure separately requires an authenticated conformity document for the import or official-release file.
Keep quality assessment and the regulatory-document gap as separate records. Follow the current authority route to obtain or clarify the document; do not turn an internal conclusion into official release.
Current authority procedure and version; document checklist; authenticated certificate where required; CoA and internal disposition; shipment identifiers; controlled correspondence.
Regulatory Affairs or the authorised import-release function, with Quality informed of any impact on status or storage.
Do not claim that a passing CoA waives a separate authority document, or that the missing document proves poor batch quality.
No matching case. Try another term.
These sources support the analytical-certificate and supplier-control checks. They do not create a universal definition or acceptance rule for every document called a CoC.
Final guidance · September 2016 · API manufacturing scope
Q7 addresses certificates of analysis, supplier evaluation and circumstances in which reliance on supplier testing may be considered.
FDA Q&A · US drug CGMP scope · contains nonbinding recommendations
The page explains representative receipt testing, identity confirmation and the role of supplier reliability under 21 CFR 211.84.
Sources reviewed 20 September 2026. Always check the current national procedure and product-specific requirement.